Cardiology products

I have an AT 2/AT 2 plus, how do I check the memory?

Hold the “Alt” key while pressing “M”.

I have an AT 2/AT 2 Plus, how do I erase the memory?

Hold the “Alt” key while pressing “E.

I have an AT 2 Plus with Spirometry, how do I get the predicted values on the Spirometer?

Typically this occurs when the entire patient data has not been entered i.e. Height, weight, sex, race, DOB

Why am I getting a lead off alarm on my AT 102?

Enter “Alt 0-3-3, this displays lead continuity which needs to be +/- 100mv.

Can I change the backlight on the LCD?

No, the back light can not be exchanged, the complete LCD assembly must be replaced.

Why is the treadmill belt not moving?

The treadmill should be connected to port 2.

Why is the treadmill belt not moving?

Check to see if the correct treadmill is selected <Exercise>, <Exercise settings>, <Treadmill/bicycle>, Select TM-400S, and Save.

I have an AT 10 with and SP 10 Spirometer, how do I get the predicted values on the Spirometer?

Typically this occurs when the entire patient data has not been entered i.e. Height, weight, sex, race, DOB

I have an AT 102 with Spirometry, how do I get the predicted values on the Spirometer?

Typically this occurs when the entire patient data has not been entered i.e. Height, weight, sex, race, DOB

I have an AT 2/AT 2 Plus, I can not get the unit to print in the auto mode but manual print works fine.

The paper was loaded incorrectly. Reload the paper with the black timing block located in the upper left corner.

I have an AT 2/AT 2 Plus, why am I getting a lead off alarm?

Enter “Alt 0-3-3, this displays lead continuity which needs to be +/- 100mv.

How many pounds can the TM-400 treadmill support?
350 pounds at full elevation and 400 pounds by maximum 3% grade
Which FAQ
We endeavour to fill this section as quickly as possible with all frequently asked questions regarding our products and services
Why am I seeing a "leads off" alarm on my AT 1
Press the "Alt" key then press 033. This displays the lead continuity, the values should be +/- 100mv.
With the AT 1 Smartprint, does the time and date print out on the internal printer?
No, the date and time only prints on the external printer. 