Hold the “Alt” key while pressing “M”.
Hold the “Alt” key while pressing “E.
Typically this occurs when the entire patient data has not been entered i.e. Height, weight, sex, race, DOB
Enter “Alt 0-3-3, this displays lead continuity which needs to be +/- 100mv.
No, the back light can not be exchanged, the complete LCD assembly must be replaced.
The treadmill should be connected to port 2.
Check to see if the correct treadmill is selected <Exercise>, <Exercise settings>, <Treadmill/bicycle>, Select TM-400S, and Save.
Typically this occurs when the entire patient data has not been entered i.e. Height, weight, sex, race, DOB
Typically this occurs when the entire patient data has not been entered i.e. Height, weight, sex, race, DOB
The paper was loaded incorrectly. Reload the paper with the black timing block located in the upper left corner.
Enter “Alt 0-3-3, this displays lead continuity which needs to be +/- 100mv.